Fracture Resistance of Single-Unit Implant-Supported Crowns: Effects of Prosthetic Design and Restorative Material


Donmez M. B., Diken Turksayar A. A., Olcay E. O., Sahmali S. M.

Journal of Prosthodontics, cilt.31, sa.4, ss.348-355, 2022 (SCI-Expanded) identifier identifier

  • Yayın Türü: Makale / Tam Makale
  • Cilt numarası: 31 Sayı: 4
  • Basım Tarihi: 2022
  • Doi Numarası: 10.1111/jopr.13415
  • Dergi Adı: Journal of Prosthodontics
  • Derginin Tarandığı İndeksler: Science Citation Index Expanded (SCI-EXPANDED), Scopus, EMBASE, MEDLINE
  • Sayfa Sayıları: ss.348-355
  • Anahtar Kelimeler: Fracture resistance, implant, PEEK, screw-cement crown
  • İstanbul Yeni Yüzyıl Üniversitesi Adresli: Hayır

Özet

Purpose: To evaluate the fracture resistance and fracture patterns of single implant-supported crowns with different prosthetic designs and materials. Materials and Methods: One hundred and forty-four identical crowns were fabricated from zirconia-reinforced lithium silicate (ZLS), leucite-based (LGC), and lithium disilicate (LDS) glass-ceramics, reinforced composite (RC), translucent zirconia (ZR), and ceramic-reinforced polyetheretherketone (P). These crowns were divided into 3 subgroups according to restoration design: cementable crowns on a prefabricated titanium abutment, cement-retained crown on a zirconia-titanium base abutment, and screw-cement crown (n = 8). After adhesive cementation, restorations were subjected to thermal-cycling and loaded until fracture. The fracture patterns were evaluated under a stereomicroscope. Statistical analysis was performed by using 2-way ANOVA/Bonferroni multiple comparison post hoc test (α = 0.05). Results: For each prosthetic design, ZR presented the highest fracture resistance (p ≤ 0.005). Other than the differences with ZLS and RC for screw-cement crowns (p > 0.05) and RC for crowns on zirconia-titanium base abutments (p > 0.05), LGC showed the lowest fracture resistance. P endured higher loads than LDS (p < 0.001), except for the crowns on zirconia-titanium base abutments (p > 0.05). Cementable crowns presented the highest fracture resistance (p < 0.001), other than LGC and LDS. The differences between LGC crowns (p > 0.05) or LDS crowns on prefabricated titanium and zirconia-titanium abutments were nonsignificant (p = 0.133). Fragmented crown fracture was predominant in most of the restorations. Screw and abutment fractures were observed in ZR screw-cement crowns, and all P crowns were separated from the abutments. Conclusions: Restorative material and restoration design affect the fracture resistance and fracture pattern of implant-supported single-unit restorations. Clinicians may restore single-unit implants in premolar sites with the materials and prosthetic designs tested in the present study.